Oil in Water | Part III | |
|
|
|
|
|
|
|
|
|
The reference voltage is checked by the ADC every ten minutes.
This indicates one or more of the following potential failure conditions:
1.The lamp has failed.
2.The cell windows have become so dirty that no light can pass through the sample.
3.The sample is opaque, due to the presence of solids or undesirable optical phenomena in the sample.
4.Condensation has occurred on the cell windows.
5.The signal processing electronics, phototube, signal cable, or demodu- lator switch has failed.
6.The filter wheel is loose on the motor shaft.
7.The reference optical filter is dirty or defective.
8.The optical bench or lamp is out of alignment.
4.0INSTALLATION
The
(50oC),(for salt water applications, the freezing point is lower due to the salt content). Cubicle installed systems (Models 6600), subject to the preceding conditions, may be installed outdoors.
Note: Ambient temperatures below 0oC, requires heated enclosures.
Only in one case when the sample system uses gravity flow should the system be installed on a level surface otherwise all other systems ar5e to be with sufficient space allocated on either side (6 feet minimum) for person- nel and test equipment access. Subject to the foregoing, the system should be placed as close to the sample point as possible and bolted to its supporting