3 Operation

Model 6600

 

 

 

 

 

 

 

 

 

3.3.5 The Model Screen

Enter the System menu, select more and Enter. The second screen ap- pears. Select more again and Enter. In the third screen select MODEL. With MODEL blinking, Enter. The screen displays the manufacturer, model, and software version information. Escape to return to the System menu.

3.3.6 Checking Linearity with ALGORITHM

From the System Function screen, select ALGORITHM, and Enter.

sel rng to set algo:

—> Ø1 Ø2 Ø3

<—

Use the UP/DOWN switch to select the range: 01, 02, or 03. Then Enter. (Some ranges may not be available, depending on your application, but at least range 01 should be).

Gas Use: SO2

Range: Ø — 10%

Enteragain.

Algorithm setup:

VERIFY SET UP

Select and EnterVERIFY to check whether the linearization has been accomplishedsatisfactorily.

Dpt

INPUT

OUTPUT

Ø

Ø.ØØ

Ø.ØØ

The leftmost digit (under Dpt) is the number of the data point being moni- tored. Use the UP/DOWN switch to select the successive points.

The INPUT value is the input to the linearizer. It is the simulated output of the analyzer. You do not need to actually flow gas.

The OUTPUT value is the output of the linearizer. It should be the ACTUAL concentrationofthespangasbeingsimulated.

If the OUTPUT value shown is not correct, the linearization must be correct- ed. Press ESCAPE to return to the previous screen. Select and Enter SET UP to Calibration Mode screen. (set-up will not work without a PC being connected totheanalyzer)

Select algorithm

mode : AUTO

3-10 Part I

Teledyne Analytical Instruments

Page 34
Image 34