Oil in Water Analyzer | Maintenance 3 | |
|
|
|
|
|
|
|
|
|
2.Check the UV source.
NOTE: Be sure to wear UV filtering eye goggles.
3.Check the solenoid valves.
3.5Service Procedures and Adjustments
3.5.1 Electronics
TAI aligns the system’s electronics. However, you may need to touch up the circuitry, using the following procedure.
Equipment Required:
Oscilloscope (dual trace is preferred, but not required) To observe oscilloscope test points switch the vertical input selector of the scope to DC.
Switch to AC to observe the demodulator switch signals.
DVM (Digital Voltmeter)
3.5.2 Power Supply Test Points
Measure +15 volt ±1 volt DC and
3.5.3Setup of the Signal Processing Front-End Amplifiers
Fill the sample cell with air or a stable fluid, such that the photo energy that strikes the detector is constant. A stable fluid is distilled or tap water. This step may be omitted when the system is stable in its present state.
Teledyne Analytical Instruments | Part II |