Oil in Water Analyzer | Operation 3 | |
|
|
|
|
|
|
|
|
|
•APPLICATION: Used to define the analysis ranges and application (gas used).
•MODEL: Displays model number and software version.
•OUTPUT_CAL:
The hierarchy of the system menu is shown in figure
3.3.1 Setting up an AUTO-CAL
Whenproperautomaticvalvingisconnected,theAnalyzercancycleitself through a sequence of steps that automatically zero and span the instrument.
Note: Before setting up an
Note: If you require highly accurate
To setup an
Choose Systemfromthemainmanu.TheVFDwilldisplayfivesubfunc-
tions.
DIG_FILT
PWD LOGOUT MORE
Select MORE and Enter
AUTOCAL HMGNZR HOLD
Use UP/DOWN to blink
ZERO | in | Ød | Øh | off |
SPAN | in | Ød | Øh | off |
Use UP/DOWN to blink ZERO (or SPAN), then Enter again. (You won’t
be able to set OFF to ON if a zero interval is entered.) A Span Every ...
(or Zero Every ...) screen appears.
Zero schedule: OFF
Day: Ød Hour: Øh