3 Operation

Model 6600

 

 

 

 

 

 

 

 

 

14.5

25.0

35.0

45.5

57.0

69.0

714.0

825.0

946.0

1090.0

At a setting of “zero”, the response time is purely set by the electronics to

4.5seconds. The numbers above can and will change depending on application andtheymerelyservetoillustratetheeffectofthedigitalfilter.

3.3.8 Homogenizer Function Setup

Dependingontheapplication,the6020samplingsystemmayhavean Ultra Sonic Homogenizer. The function of this part is to prevent the oil in the waterfromclumpingtogether.Thehomogenizershouldturnonaftertheinitial warm up and self-diagnostic period when the analyzer enters the Analyze mode. Thehomogenizerwillturnoffautomaticallyassoonastheanalyzerentersthe Zero and Span mode, turning back on at the end.

Undersomeconditions,itmightbedesirabletomanuallyturntheUltra SonicHomogenizeroff. Thehomogenizersetupfunctionisprovidedinthe Systemmenu.ToaccesstheHomogenizerfunctionsetup:

1.EntertheSystemmenu.

2.Select MORE in the first System menu using the UP/DOWN switch.

3.Select HMGNZR in the second System menu, the screen will display

Set Ultra Sonic

Homogenizer: ON

4.By using the UP/DOWN switch the homogenizer can be toggled on and off.

5.Enter or Escape to return to the analyze mode.

3-12 Part I

Teledyne Analytical Instruments

Page 36
Image 36
Teledyne 6600 manual Homogenizer Function Setup, Set Ultra Sonic Homogenizer on