3 Operation | Model 6600 | |
|
|
|
|
|
|
|
|
|
3.3.10 Calibration/Hold Timer Setup
ThisCalibrationTimerletstheoperatoradjustthetimetheinstrument purgesthecalibrationgaspriortoactuallystartingthecalibrationcomputations. The Sample timer lets the operator adjust the time the instrument purges sample gasafterfinishingacalibrationbeforeitletstheanalogoutputsanddisplaytrack thechangeinconcentration.
ThisfunctionandtheTRACK/HOLDfeaturewillpreventfalsealarms whileperformingremoteorautoscheduledcalibrations. Thesefunctionsarenot applicableifthecalibrationisinitiatedthroughthefrontpanel. Toenterthe Calibration/HoldTimerfunction,youmust:
1.EntertheSystemmenu:
DIG_FILT
PWD LOGOUT MORE
2.Using theUP/DOWN switch, select MORE and press Enter: The Second System screen appears:
AUTOCAL HMGNZR TRACK
or
AUTOCAL HMGNZR HOLD
3.Select with the UP/DOWN switch
Calbrt hold: 3 min
Sample hold: 1 min
The calibration hold time is set on the first row, while the sample hold time is set on the second row. To select one or the other, use the Right or Left keys. To modify the time of either timer, use the Up or Down keys. The time is in the minutes.
3.3.11 Analog 4-20mA Output Calibartion
This function will let the operator calibrate the 4 to 20 mA analog output to match the display reading. A DMM configure as a DC ammeter is needed. The DMM should be connected across the output terminals of the 4 to 20 mA output to monitor the output current. To enter the 4 to 20 mA output adjust function, youmust:
1.EntertheSystemmenu: