Oil in Water Analyzer | Part I: Control Section | |
|
|
|
|
|
|
|
|
|
Table
Range | Voltage (V) | Current (mA) |
Range 1 | 0.25 | 8 |
Range 2 | 0.50 | 12 |
Range 3 | 0.75 | 16 |
Alarm Relays:
There are three
(under RELAY OUTPUTS) for making connections to internal alarm relay contacts. Each provides a set of Form C contacts for each type of alarm. Each has both normally open and normally closed contact connections. The contact connections are indicated by diagrams on the rear panel. They are capable of switching up to 3 ampers at 250 V AC into a resistive load (Figure
Normally closed |
Normally open |
Moving contact |
Normally open |
Moving contact |
Figure 2-6: Types of Relay Contacts
The connectors are:
Threshold Alarm 1: • Can be configured as high (actuates when concentration is above threshold), or low (actuates when concentration is below thresh old).
•Can be configured as
•Can be configured as latching or nonlatching.
•Can be configured out (defeated).
Teledyne Analytical Instruments | Part I: |