Oil in Water Analyzer

Part I: Control Section

 

 

 

 

 

 

 

 

 

Fuse Installation: The fuse holders accept 5 x 20 mm, 4.0 A, T type (slow blow) fuses. Fuses are not installed at the factory. Be sure to install the proper fuse as part of installation (See Fuse Replacement in chapter 4, maintenance.)

Analog Outputs: There are eight DC output signal connectors on the ANALOG OUTPUTS terminal block. There are two connectors per output with the polarity noted. See Figure 2-5.

The outputs are:

0–1 V dc % of Range: Voltage rises linearly with increasing sample con- centration, from 0 V at 0% to 1 V at 100%. (Full scale = 100% programmed range.)

0–1 V dc Range ID: 0.25 V = Range 1, 0.5 V = Range 2, 0.75 V = Range 3.

4–20 mA dc % Range: (-M Option) Current increases linearly with increas- ing sample concentration, from 4 mA at 0% to 20 mA at full scale 100%. (Full scale = 100% of programmed range.)

4–20 mA dc Range ID: (-M Option) 8 mA = Range 1, 12 mA = Range 2, 16 mA = Range 3.

Figure 2-5: Analog Output Connections

Examples:

Teledyne Analytical Instruments

Part I: 2-3

Page 17
Image 17
Teledyne 6600 manual Examples, Analog Output Connections