Oil in Water Analyzer

Operation 3

 

 

 

 

 

 

 

 

 

Protected until password entered

After two seconds it will return to the System menu.

3.3.4 System Self-Diagnostic Test

The Model 6600 has a built-in self-diagnostic testing routine. Pre-pro- grammed signals are sent through the power supply, output board, preamp board and sensor circuit. The return signal is analyzed, and at the end of the test the status of each function is displayed on the screen, either as OK or as a number between 1 and 1024. (See System Self Diagnostic Test in chapter 5 for number code.)

Note: The sensor will always show failed unless Zero fluid is present in the sampling cell at the time of the SELF-TESTinput thru the sample inlet.

Theselfdiagnosticsarerunautomaticallybytheanalyzerwheneverthe instrument is turned on, but the test can also be run by the operator at will. If any of the functions fails, the System Alarm is tripped, but is not tripped at the startup because it might give false alarm after a power failure. To initiate a self diagnostictestduringoperationentertheSystemmenu.

DIG_FILT SELF-TEST

PWD LOGOUT MORE

Use the UP/DOWN again to move the blinking to the SELF–TESTand Enter.Thescreenwillfollowtherunningofthediagnostic.

RUNNING DIAGNOSTIC

Testing Preamp — Cell

When the testing is complete, the results are displayed.

Power: OK Analog: OK

Cell: 2 Preamp: 3

The module is functioning properly if it is followed byOK.A number indicates a problem in a specific area of the instrument. Refer to Chapter 5 Maintenance and Troubleshooting for number-code information. The results screenalternatesforatimewith:

Press Any Key

To Continue...

ThentheanalyzerreturnstotheinitialSystemscreen.

Teledyne Analytical Instruments

Part I 3-9

Page 33
Image 33
Teledyne 6600 manual System Self-Diagnostic Test