Oil in Water Analyzer

Maintenance 3

 

 

 

 

 

 

 

 

 

3.5.5Balancing the Optics for Equal Light Transmission with Zero Fluid in the SAMPLE CELL

The objective of this procedure is to obtain measuring and reference peak heights as displayed on the oscilloscope that are approximately equal, with the tallest peaks set at 8 to 9 volts. This must be done with zero fluid in the cell. (Collect a Zero prepared fluid from the sample system ifor all oil in water analyzers). See Part III, Section 5.5

The procedure is purely mechanical and consists of adjusting the amount of light passing through either the measuring or reference filter, never both. Screens (wire mesh) of varying density are used for this operation and are part of the small took kit accompanying the instrument.

1.Observe the oscilloscope and judge if optical balancing is needed. When the difference is less than 1 volt, balancing is not required. The tallest of the two peaks should be adjusted to 8 or 9 volts with the gain control R2 on the detector PC board. When this cannot be done because both peaks are too short or too long, search for screens mounted in the light path, usually located in a holder on the light pipe which interconnects the detector and sample module, and remove or add screens, as necessary.

2.When balancing is needed, identify the peaks as outlined under Section

3.For example, if the reference peak is the shorter one, stop the filter wheel with your hand and see if screens are located behind the reference filter. The reference filter is identified by the letter “R” engraved on the filter wheel.

4.If screens are found, remove them after taking the filter wheel off the shaft with the special Allen wrench supplied in the tool kit.

5.After removal of the screens and remounting the filter, mount the filter wheel back on the shaft. Position it correctly on the shaft by lining up the two paint marks on shaft and wheel.

6.Turn on the instrument and observe the balance on the oscillo- scope.

a.If the reference peak is now too tall, remove the filter wheel and add a screen of lesser density behind the reference filter. Repeat this procedure until the peaks are within 1 volt of each other.

Teledyne Analytical Instruments