drawings)

Part III

Model 6600

 

 

 

 

 

 

 

 

 

c Check to see that lines are properly heat traced if required for ambient temperature extremes and that the sample will not boil from being too hot (for example: (Maximum temperature allowed is:52 degrees C (limited by the plastic filter housings), Internal sample system operation Pressure limit is:150psig G for produced water). The inlet sample pressure (457 psig) is reduced by a forward pressure regulator (PR1) which is followed by a protecting relief valve in the event of failure of PR1).

d Check and familiarize oneself of the analyzer and sample system components and to assure the entire piping flow agrees with the Piping Drawing (see addendum

drawings)

1Sample connections; Refer to Piping diagram

(see addendum drawings)

a inlet-

-sample enters the pump.

breturns

- sample flows through Cell then exits the system to a common return

cdrain

Assure the common drain(s) is always free flowing and always at atmospheric pressure (unless back to process by suitable differential or pumping) and will never freeze or block (backup) under icy conditions.

dGrab Sample port: Refer to your particular piping scheme for obtaining the sample for analysis,etc.OBSERVEEXTREME CAUTIONwhen operating any needle

1-52 Part III (see addendum drawings) Teledyne Analytical Instruments