2 Installation | Model 6600 | |
|
|
|
|
|
|
|
|
|
The span fluid must be the major component of the sample mixed with a small amount of the component of interest. The concentration must be 60 to 80% of the range or the widest range of the instrument (if the instrument provides more than one range).
2.4.2 Calibration
Refer to Section 3.3.8 section I of the manual to determine how to manipulate the mode setting. Two calibration methods are available.
1.Calibration with zero and span fluids.
2.Calibration with a span filter.
Method One:
1.Inject zero fluid and set zero as referred in Part I
2.Inject span fluid and set the concentration of the span fluid with the span procedure referred in Part I
Method Two:
1.Determine the span setting using Method One.
2.Activate the span filter (as referred in section 3.3.8) Part I
3.Record the display reading (this is the span filter reading and must be recorded).
4.You can calibrate the instrument now with the span filter.
Power up the system, and test it as follows: 1. Repeat the
| Teledyne Analytical Instruments |