3 Operation

Model 6600

 

 

 

 

 

 

 

 

 

Anlz indicates that the analyzer is in the Analyze mode. If there is an * before the Anlz, it indicates that the range is linearized.

1.95

ppm

SO2

R1:Ø

—10

*Anlz

If the concentration detected is overrange, the first line of the display blinks continuously.

3.8Programming

CAUTION: The programming functions of the Set Range and Curve Algorithm screens are configured at the facto- ry to the users application specification. These func- tions should only be reprogrammed by trained, qualified personnel.

Toprogram,youmust:

1.Enter the password, if you are using the analyzer’s password protection capability.

2.Connect a computer or computer terminal capable of sending an RS-232 signal to the analyzer RS-232 connector. (See chapter 2 Installation for details). Send the rp command to the analyzer.

3.Enter the System menu.

DIG_FILT SELF-TEST

PWD LOGOUT MORE

Use the UP/DOWN switch to blink MORE, then Enter.

AUTOCAL HMGNZR HOLD

CAL-HOLD-TIMER MORE

Select MORE and ENTER one more time

ALGORITHM APPLICATION

MODEL OUTPUT: 4MA

Now you will be able to select the APPLICATION and ALGORITHM set-upfunctions.

3-28 Part I

Teledyne Analytical Instruments

Page 52