Appendix | Models 6600 | |
|
|
|
|
|
|
|
|
|
Typical Analytical Performance Specifications:
(will vary per application)
Accuracy: | ±2% of full scale possible (Oil in Water) |
| When calibrated on specific oil of interest. |
Noise: | Less than ±1% |
Drift: | Less than ±1% per auto Zero cycle (source/ |
| detector dependent) |
Sample Cell: | (Aluminum/CPVC for Oil in Water) with |
| Sapphire/Quartz windows standard. (Kynar, |
| Kalrez optional) |
Cell Length: | 1/8 to 3.5” (application dependent) |
Flow Rate: | 100 ccm to 2 lpm (application dependent) |
Light Source: | Mercury, Hg |
Sensitivity: | .08 to 1.0 absorbance units. |
Reproducibility: | |
Filter Wavelength: | 210 to 1000 millimicrons. (application depen- |
| dant - oil in water 254/365mm). |
Sample Pressure: | 150 psi maximum (oil in water applications) |
Sample temperature: |
Teledyne Analytical Instruments |