Teledyne 6600 Typical Analytical Performance Specifications, Appendix, Models, Accuracy, Noise

Models: 6600

1 151
Download 151 pages 30.8 Kb
Page 149
Image 149
Models 6600

Appendix

Models 6600

 

 

 

 

 

 

 

 

 

Typical Analytical Performance Specifications:

(will vary per application)

Accuracy:

±2% of full scale possible (Oil in Water)

 

When calibrated on specific oil of interest.

Noise:

Less than ±1%

Drift:

Less than ±1% per auto Zero cycle (source/

 

detector dependent)

Sample Cell:

(Aluminum/CPVC for Oil in Water) with

 

Sapphire/Quartz windows standard. (Kynar,

 

Kalrez optional)

Cell Length:

1/8 to 3.5” (application dependent)

Flow Rate:

100 ccm to 2 lpm (application dependent)

Light Source:

Mercury, Hg

Sensitivity:

.08 to 1.0 absorbance units.

Reproducibility:

+/-1% of scale or better

Filter Wavelength:

210 to 1000 millimicrons. (application depen-

 

dant - oil in water 254/365mm).

Sample Pressure:

150 psi maximum (oil in water applications)

Sample temperature:

1-120oC (34-250oF) (non freezing, non-boiling)

A-2

Teledyne Analytical Instruments

Page 149
Image 149
Teledyne 6600 Typical Analytical Performance Specifications, Appendix, Models, will vary per application, Accuracy, Noise