Part I: Control Section | Maintenance 4 | |
|
|
|
|
|
|
|
|
|
Preamp
0OK
1Zero too high
2Amplifier output doesn't match test input
3Both Failed
>3 Call factory for information
Detector
0OK
1Failed (open filament, short to ground, no power.)
2Unbalance (deterioration of filaments, blocked tube)
4.3Major Internal Components
All internal components are accessed by unbolting and swinging open the front cover, as described earlier. The major internal component locations are shown in Figure
The 6600 contains the following major internal components:
•Customer Interface PCB (Power Supply on bottom surface)
•Preamp PCB (Contains Microprocessor)
•Front Panel PCB (Contains Displays)
5 digit LED meter
2 line, 20 character, alphanumeric, VFD display
See the drawings in the Drawings section in back of this manual for details.
For Optical/Detector Alignments, refer to parts II or III of this manual
Teledyne Analytical Instruments | Part I: |