Part I: Control Section | Maintenance 4 | |
|
|
|
|
|
|
|
|
|
Maintenance
Aside from normal cleaning and checking for leaks at the gas connec- tions, routine maintenance is limited to replacing filter elements and fuses, and recalibration.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS MANUAL.
4.1Fuse Replacement
The 6600 requires two 5 x 20 mm, 4 A, T type (Slow Blow) fuses.
The fuses are located inside the main housing on the Electrical
Connector Panel, as shown in Figure
1.Disconnect the Unit from its power source.
2.Place a small screwdriver in the notch in the fuse holder cap, push in, and rotate 1/4 turn. The cap will pop out a few millimeters. Pull out the fuse cap and fuse, as shown in Figure
Teledyne Analytical Instruments | Part I: |