Part III | Model 6600 | |
|
|
|
|
|
|
|
|
|
The system safe vent should be equipped to accept a
5.0SYSTEM START-UP/CALIBRATION
The information contained in the following subsections deals primarily with the steps necessary before the total system can be used for continuously monitoring oil in water. This information involves installation checks, electrical checks, fluid dynamics, and calibration.
Due to variations in water and oils as encountered under field conditions, certain adjustments must be made to optimize system performance for the specific site. This includes calibration.
5.1Installation Check
1.Observe that power, signal and alarm connections are properly made and that the system is properly grounded. Refer to Installation Section of manual and to Outline and Interconnection Drawings.
2.Inspect the external sample delivery system, including sample pump (when applicable), sample take up point, and source of water sample.
3.Confirm that the recorder is of the correct type.
5.2Electronics Check
1.Check that all PC boards are in place.
2.Turn off ultrasonic homogenizer by rotating the homogenizer potenti- ometer fully
3.Switch on electronics and confirm that 15 volt, and
4.With an oscilloscope, check test points and reed switch action with air or