3 Operation

Model 6600

 

 

 

 

 

 

 

 

 

-Enter the System Menu

-Select MORE on the first and second System menu screens. -In the last System Menu screen you will see:

ALGORITHMAPPLCATION

MODEL OUT_CAL ANLZ

-Selectthelastfield“ANLZ”usingtheUp/Downswitch.

-Press Enter to change the mode of the filter and the solenoid. The se- quenceisasfollows:

1.ZERO:Sets the Filter and solenoid in the zero mode (span filter off, zerosolenoidon).

2.SPN1: Sets the Filter and solenoid in the span mode (span filter on, zerosolenoidon).

3.SPN2:Sets second span filter on (usually not installed) and zero solenoidon

4.SPNB:Setsboth span filter on(usuallyonly onefilter installed) and zero solenoid on.

5.ANLZ:Returns Filtersand solenoid totheAnalyze mode (spanfilter off,zerosolenoidoff).

-Press Escape to see the effect

NOTE: FOR PROPER OPERATION OF THE ANALYZER

RETURN TO ANLZ MODE.

3.4The Zero and Span Functions

The Model 6600 can have as many as three analysis ranges plus a special calibrationrange(CalRange).Calibratinganyoneoftherangeswillautomatical- ly calibrate the other ranges.

CAUTION: Always allow one hour warm-up time before calibrat- ing, if your analyzer has been disconnected from its power source. This does not apply if the analyzer

was plugged in but was in STANDBY.

The analyzer is calibrated using zero, and span gases.

3-16 Part I

Teledyne Analytical Instruments

Page 40
Image 40