Part III | Model 6600 | |
|
|
|
|
|
|
|
|
|
4Turn on utility power using circuit breaker switch to systems.
5Turn on bypass pump when applicable, and assure
7 Once assured sample inlet is proceed as follows:
aSwitch on V1 using the sample/utility valve set to utility position with the same V1 open Observe the cleanliness of the utility water to the drain. (You can again also check water quality at the sample grab point).
If clear, then turn on Sample pump using sample pump power (which also turns on the homogenizer when applicable (application dependent) Allow several minutes for the sample to flush through the entire measuring train of the system to drain or safe vent.
8 Allow warm-up for 1 hour
The UV lamp in particular requires the most time to stabilize from a cold start. If the unit has been turned of temporarily for only a few minutes,
C Testing the system
A Control Unit
1Perform a
2Refer to Section 2.3, Part I, for Testing of System on Control Unit.
3Please refer to Section 3.0 for Setup parameters, operation, programming of the Control Unit.
B Analysis Unit