Teledyne 6600 manual 2.3Testing the System, 2.4Calibration 2.4.1 Calibration Fluids, Part II

Models: 6600

1 151
Download 151 pages 30.8 Kb
Page 75
Image 75
2.2.7 Draining the System

Oil in Water Analyzer

Part II: Analysis Unit

 

 

 

 

 

 

 

 

 

2.2.7 Draining the System

In liquid analysis systems, the system return must terminate back to the process or a safe area as the sample may be poisonous or corrosive. Olso, the return pressure must be always sufficiently low enough from the inlet pressure to maintain proper response times within the system.

2.3Testing the System

Before plugging the instrument into the power source:

Check the integrity and accuracy of the fluid connections. Make sure there are no leaks.

Check the integrity and accuracy of the electrical connections. Make sure there are no exposed conductors

Check that sample pressure is between 3 and 40 psig, according to the requirements of your process.

NOTE: Special designed systems may require checks under vacuum or high pressure (consult manual addendum). Consult commissioning start-up section in the manual addendum.

Warning: Do not operate the “ultrasonic homogenizer” in the instrument for more than one (1) minute without a liquid sample properly flowing through the homog- enizer.

Power up the system, and test it by performing the following operations:

1. Repeat the Self-Diagnostic Test, section 3.3.4, part I

2.4Calibration

2.4.1 Calibration Fluids

Zero and span fluids must be made by the chemistry lab or certified zero and span fluids bought from a supplier. The zero fluid must be the major component of the sample, free from the component of interest.

Teledyne Analytical Instruments

Part II: 2-3

Page 75
Image 75
Teledyne 6600 manual 2.3Testing the System, 2.4Calibration 2.4.1 Calibration Fluids, Part II, Oil in Water Analyzer