3 Operation | Model 6600 | |
|
|
|
|
|
|
|
|
|
the analyzer is zeroed and spanned as usual. Then, each special calibration gas, foreachoftheintermediatecalibrationpoints,isflowed,inturn,throughthe sensor.Aseachgasflows,thedifferentialvalueforthatintermediatecalibration point is entered from the front panel of the analyzer.
Note: The span gas used to span the analyzer must be >90% of the range being analyzed.
Beforestartinglinearization,performastandardcalibration.Seesection 4.4. To enter data:
From the System Functions screen—
1.Select ALGORITHM , and Enter.
2.Select and Enter SETUP.
3.Enter AUTO from the Calibration Mode Select screen.
The Auto Linearize Mode data entry screen appears.
19.5 ppm SO2 Input(Ø) :20.00
5.Use the UP/DOWN switch to set the proper value of calibration gas, and Enter. Repeat this step for each
6.Repeat step 5 for each of the special calibration gases, from the lowest to the highest concentrations. Escape when done.
To end the session, send:
st<enter> st<enter>
to the analyzer from the computer.