
3 Operation | Model 6600 | |
|
|
|
|
|
|
|
|
|
Sel rng to set appl:
Use the UP/DOWN switch to increment/decrement the range number to 01, 02, 03, or CAL, and Enter.
Gas | Name | ********** |
FR:Ø | TO:1Ø | % |
Use the UP/DOWN switch to increment the respective parameters as desired, and Enter to move to the next. On the last field %/ppm Enter to accept the values and return to range selection menu. (See note below.) Repeat for each range you want to set.
Note: The ranges must be increasing from low to high, for example, if Range 1 is set to
Ranges, alarms, and spans are always set in either percent or ppm units, as selected by the operator, even though all
Note: When performing analysis on a fixed range, if the concentra- tion rises above the upper limit as established by the operator for that particular range, the output saturates at 1 V dc (or 20 mA). However, the digital readout and the
To end the session, send:
st<enter> st<enter>
to the analyzer from the computer.
3.8.2 The Curve Algorithm Screen
The Curve Algorithm is a linearization method. It provides from 1 to 9
intermediate points between the ZERO and SPAN values, which can be normal-
Eachrangeislinearizedindividually,asnecessary,sinceeachrangewill usuallyhaveatotallydifferentlinearizationrequirement.
To linearize the ranges, you must first perform the four steps indicated at the beginning of section 3.8 Programming. You will then be in the second System menu screen.