Probe Operation

Restarting the Probe

Selecting the Cold Start Menu Item

Use the following procedure to cold start the probe using the probe’s Main Menu:

1.Connect a local terminal (or a PC emulation a terminal) to the probe. Refer to “Probe Operation” on page 91 for information on connecting a local terminal.

2.Press the CONFIG button once (on the back of the probe) to place the probe into the configuration mode. Use a narrow, pointed object (like a pen) to press the recessed CONFIG button. The probe then displays its Main Menu on the terminal. Figure 5-1 on page 95 shows the probe’s Main Menu.

3.Press 8 to execute a cold start and to exit the probe’s Main Menu. The cold start occurs immediately. The Activity and Fault LEDs are turned on during a cold start. When the cold start completes, the Activity LED flashes to indicate traffic (if present), the Fault LED turns off, and the ~Line On (or Power) LED is on.

97