Agilent Technologies J3972A, J3919A manual Table Installation, Rack or Cabinet Installation

Models: J3919A J3972A

1 189
Download 189 pages 36.38 Kb
Page 65
Image 65

Installation

Installing the Probe

Table Installation

Use the following procedure to install the probe on a table:

1.Attach the self-adhesive rubber feet to the bottom of the probe as marked in each corner.

2.Place the probe on a flat surface (refer to the requirements as listed in “Selecting a Location” on page 43).

Rack or Cabinet Installation

You can install the probe in a rack or cabinet with either the front or rear panel facing out. You may want the rear panel facing out so that you can see the status LEDs for each port. (The front panel Fault and Activity LEDs indicate when the 10Base-T/100Base-TX telemetry interface LEDs are lit.)

Make sure you have screws that fit your rack (or cabinet) before installing the probe as noted in Step 3 below.

Use the following procedure to install the probe in a rack or cabinet:

1.Attach the installation brackets to the probe with the 10-mm #M4 screws (included), using a POZIDRIVâ #2 or cross-head screwdriver as shown in Figure 3-1 on page 46.

2.Position the probe in the rack (or cabinet) and slide it up or down until the rack holes line up with the holes in the brackets.

3.Attach the probe to the rack with the 5/8-in. #12-24 screws (included). Some cabinets require 5/8-in. #10-32 screws (available as Agilent part number 2680-0302) instead of the 5/8-in. #12-24 screws.

Figure 3-1 on page 46 shows a rack or cabinet installation.

45

Page 65
Image 65
Agilent Technologies J3972A, J3919A manual Table Installation, Rack or Cabinet Installation