1From the HOME screen, press . The SET EASY BOLUS screen will appear. The single “step” value will be flashing.

2 Press

the number of times needed for your bolus

 

amount. Watch the amount change on the screen with each

x 10

press. When your total bolus amount appears on the screen,

 

press ACT. The pump will vibrate or sound a different tone

 

for each

press. Listen/feel to count the steps without

 

looking at the screen.

SET EASY BOLUS

0.1U

(flashing)

SET EASY BOLUS

1.0U ACT

For example: You need to deliver a 1.0 unit bolus, and your step size is set to 0.1. Each time you press , the units increase by the “step” amount. To deliver 1.0 units, you need

to press the button 10 times. (10 X 0.1 = 1.0) The screen will show 1.0 units.

NOTE - Pressing or ESC will cancel the Easy bolus. For your safety you cannot use the to select the Easy Bolus values.

3If this amount is correct, press ACT to start the Easy Bolus delivery. The BOLUS DELIVERY screen will show the units being delivered. When the total bolus is finished, the pump will beep or vibrate.

If this amount is wrong, press ESC or to start over. The pump will return to the HOME screen.

SET EASY BOLUS

ACT

1.0U to accept

ESC

to cancel

BOLUS DELIVERY

1:32P 0.1U Normal

Optimizing pump therapy 93

Page 115
Image 115