Stopping your pump

You can stop your pump with the suspend function. Suspend stops all insulin delivery including the current basal and any bolus or prime deliveries that are in progress. While suspended, your pump will not deliver insulin until you Resume your pump. When basal is resumed, the pump is taken out of the Suspend mode.

The pump will beep or vibrate about every 15 minutes on the hour to remind you that it is not delivering insulin. Example: You suspend your pump at 11:20AM. The pump will beep/vibrate at 11:30AM, 11:45AM, 12:00PM, and so on until you resume your pump (basal resumes).

NOTE - When suspended, your pump is in Attention mode (a solid circle). When in Suspend, you can only resume your basal or view the STATUS screen. No other functions are available.

Do these steps to suspend your pump:

1Select Suspend from the MAIN MENU, and press ACT.

ACT

MAIN MENU

Bolus

Suspend

Basal

2SUSPEND will flash. Press ACT to stop your pump.

ACT

SUSPEND

Press ACT to Stop Pump

(flashing)

3

The screen will show that the pump is suspended and the time that it stopped. After a few minutes, the pump will default to the HOME screen with a solid circle.

Stopped at 1:05P

SUSPEND

NOTE - Press ESC to view the STATUS screen and verify your pump is suspended.

STATUS U100

Suspended at 01:05P

Last Bolus N 3.8U

7:58A 22 FEB

Basic programming 39

Page 61