Pump specifications 145
Pump motor
The pump motor has a unique, patented design with an integrated safety check system. The system delivers in
precise increments.
Dual Wave bolus
Delivers a normal bolus followed by a Square Wave bolus (limited by maximum bolus).
Easy bolus
Features programming using audible tones (or vibrate pulses) in user-defined increments. Beep mode range: 0 to
maximum bolus; Vibrate mode range: 0 to 20 steps or maximum bolus, whichever comes first.
Default step increment: 0.1 unit
Step size < maximum bolus.
Adjustable step size: 0.1 to 2.0 units per step.
Accessible from the remote control or pump buttons.
Infusion pressure
Maximum infusion pressure and occlusion pressure:
13.7 PSI (pounds per square inch)
94,46 kPa (Kilopascal)
(Insulin) sensitivity
Low resv (reservoir) warning
Values are based on displayed amount, not actual amount.
Maximum settings: factory setting: range: warning limits:
8 50 mg/dL (2.8 mmol/L) (mg/dL) 10 – 400 mg/dL
(mmol/L) 0.5 – 22.2 mmol/L less than 20 or greater than 100 mg/dL
less than 1.1 or greater than 5.6 mmol/L
time: 2 – 24-hours, and @ 1:00 hour before empty 08:00 hours (default when time is selected)
units: 5 – 50 units, and @ 1/2 amount remaining 20 units (factory default)