Pump motor

The pump motor has a unique, patented design with an integrated safety check system. The system delivers in precise increments.

Dual Wave bolus

Delivers a normal bolus followed by a Square Wave bolus (limited by maximum bolus).

Easy bolus

Features programming using audible tones (or vibrate pulses) in user-defined increments. Beep mode range: 0 to

maximum bolus; Vibrate mode range: 0 to 20 steps or maximum bolus, whichever comes first.

Default step increment:

0.1 unit

Step size

< maximum bolus.

Adjustable step size:

0.1 to 2.0 units per step.

Accessible from the remote control or pump buttons.

Infusion pressure

Maximum infusion pressure and occlusion pressure:

13.7 PSI (pounds per square inch)

94,46 kPa (Kilopascal)

(Insulin) sensitivity

Maximum settings:

factory setting:

range:

 

warning limits:

8

50 mg/dL (2.8 mmol/L)

(mg/dL)

10 – 400 mg/dL

less than 20 or greater than 100 mg/dL

 

 

(mmol/L)

0.5 – 22.2 mmol/L

less than 1.1 or greater than 5.6 mmol/L

 

 

 

 

 

Low resv (reservoir) warning

Values are based on displayed amount, not actual amount.

time:

2

24-hours, and @ 1:00 hour before empty

08:00 hours (default when time is selected)

 

 

 

 

units:

5

– 50 units, and @ 1/2 amount remaining

20 units (factory default)

 

 

 

 

Pump specifications 145

Page 167
Image 167
Energizer 515, 715 manual Dual Wave bolus