Alarm conditions
Alarms put the pump in “Attention” mode.
A (Alarm)
This alarm shows an “A” followed by two numbers.
Auto off
Alerts you that no buttons were pressed during the time limit you set for the AUTO OFF DURATION feature, and so insulin delivery has been stopped.
9:42AAxx Pump reset
Settings preserved ESC, ACT to clear
XXindicates the alarm number
11:17AAUTO
Delivery stopped No OFF buttons pushed
during time limit ESC, ACT to resume delivery
Batt out limit
Occurs if the battery has been out of the pump for more than five minutes. Verify that the pump time and date are correct. If not correct, go to the Utilities Menu and reset the date and time.
Bolus stopped
If this alarm occurs, it is very important to check bolus history to review how much of the bolus was actually delivered. If necessary, reprogram a bolus with the amount not yet delivered.
12:00ABATT
OUT LIMIT
Batt change too slow
ESC, ACT to clear
8:35A BOLUS
STOPPED
Loose battery cap? Pump dropped or
bumped?
Check bolus history, Reprogram bolus
if required ESC, ACT to clear