Troubleshooting 137

Alarm conditions
Alarms put the pump in “Attention” mode.

A (Alarm)

This alarm shows an “A” followed by two numbers. A-alarms cause all
insulin delivery to stop. Your pump settings are retained. If this alarm
repeats often, call the Medtronic MiniMed 24-hour Product Help Line for
assistance.

Auto off

Alerts you that no buttons were pressed during the time limit you set
for the AUTO OFF DURATION feature, and so insulin delivery has been
stopped.

Batt out limit

Occurs if the battery has been out of the pump for more than five
minutes. Verify that the pump time and date are correct. If not correct, go
to the Utilities Menu and reset the date and time.

Bolus stopped

If this alarm occurs, it is very important to check bolus history to review
how much of the bolus was actually delivered. If necessary, reprogram a
bolus with the amount not yet delivered.

Axx

9:42A
Pump reset
Settings preserved
ESC, ACT to clear
XX indicates the
alarm number

AUTO

OFF

Delivery stopped No
buttons pushed
during time limit
ESC, ACT to resume
delivery
11:17A
Batt change
too slow
ESC, ACT to clear
12:00A

OUT LIMIT

BATT

8:35A BOLUS

STOPPED

Loose battery cap?
Pump dropped or
bumped?
Check bolus history,
Reprogram bolus
if required
ESC, ACT to clear