Alarm conditions

Alarms put the pump in “Attention” mode.

A (Alarm)

This alarm shows an “A” followed by two numbers. A-alarms cause all insulin delivery to stop. Your pump settings are retained. If this alarm repeats often, call the Medtronic MiniMed 24-hour Product Help Line for assistance.

Auto off

Alerts you that no buttons were pressed during the time limit you set for the AUTO OFF DURATION feature, and so insulin delivery has been stopped.

9:42AAxx Pump reset

Settings preserved ESC, ACT to clear

XXindicates the alarm number

11:17AAUTO

Delivery stopped No OFF buttons pushed

during time limit ESC, ACT to resume delivery

Batt out limit

Occurs if the battery has been out of the pump for more than five minutes. Verify that the pump time and date are correct. If not correct, go to the Utilities Menu and reset the date and time.

Bolus stopped

If this alarm occurs, it is very important to check bolus history to review how much of the bolus was actually delivered. If necessary, reprogram a bolus with the amount not yet delivered.

12:00ABATT

OUT LIMIT

Batt change too slow

ESC, ACT to clear

8:35A BOLUS

STOPPED

Loose battery cap? Pump dropped or

bumped?

Check bolus history, Reprogram bolus

if required ESC, ACT to clear