Inserting the reservoir in your pump
If your reservoir is already inserted in your pump, continue to “Manual prime” on page 49.
You must do these steps in the order described. Your pump screen will show instructions to help you with these steps. If you are practicing, do NOT insert the reservoir in your pump.
CAUTION: You must rewind your pump before installing a new reservoir. As part of the pump’s function, it measures the reservoir volume. To ensure correct volume measurements, your pump has been designed to require a rewind before you insert your reservoir.
1If you are using the pump for the first time, remove the shipping cap from the reservoir compartment.
WARNING: Do not insert the reservoir in the pump if you did not rewind. Doing so could result in inaccurate insulin delivery.
Never insert the reservoir into the pump while the tubing is connected to your body. Doing so could result in an accidental infusion of insulin.
2Insert the reservoir into the top of the pump case.
3Turn the tubing connector approximately
tubing connector
ESC
ACT
4Attach the activity guard, if desired.
5You must now do a manual prime as described in the next section.