102 Chapter: 6

NOTE - The pump delivers basal amounts in 0.05 U/H increments. Because of this, your temp basal

amount will be rounded down to the next 0.05 U/H increment.

temp basal start time
(120% x rate 3 = 2.34)
2.34 rounded down to
the next 0.05 increment
is 2.30 U/H
1.50 U/H
(rate 1)
2.55 U/H
(rate 2) 1.95 U/H
(rate 3)
3.05
U/H 2.30
U/H
(120% x rate 2 = 3.06)
3.06 rounded down to
the next 0.05 increment
is 3.05 U/H
temp basal stop time
temp basal settings
temp basal type: Percent of basal
duration: 4 hours (1:05p - 5:05p)
rate: 120 percent (%)
5:05 p.m.
(17:05)
7:00a.m.
(07:00) 1:05 p.m.
(13:05)
3:00 p.m.
(15:00) 12:00 a.m.
(00:00)