#3: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a blood glucose of 70.

 

(food estimate)

 

 

60g

= 2 units

+

 

 

 

 

30g/u

 

 

 

 

= 2 + (-0.5) = 1.5 unit

estimate = 1.5 unit

 

(correction)

 

 

70mg/dL - 90mg/dL

=

-20mg/dL

= -0.5 unit

 

 

 

 

 

 

40mg/dL/u

40mg/dL

 

 

 

 

 

 

 

#4: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a blood glucose of 100.

 

(food estimate)

 

 

60g

= 2 units

+

 

 

 

 

30g/u

 

 

 

 

= 2 + 0 = 2 units

estimate = 2 units

(correction)

Correction is 0 because the current blood glucose reading is between the BG High and Low targets

#5: Previous bolus activity results in a calculation of 1.5 units unabsorbed (active) insulin. User inputs 60 grams of carbs and a blood glucose of 200.

(food estimate)

60g

30g/u = 2 units

+

(correction)

200mg/dL - 120mg/dL - 1.5 units (active insulin) = 2 - 1.5 = 0.5 units

40mg/dL/u

=2 + 0.5

=2.5 units estimate = 2.5 units

Pump specifications 153

Page 175
Image 175
Energizer 515, 715 manual Correction 70mg/dL 90mg/dL 20mg/dL