#6: Previous bolus activity results in a calculation of 3.5 units unabsorbed (active) insulin. User inputs 60 grams of carbs and a blood glucose of 200.

(food estimate)

 

 

60g

= 2 units

+

 

 

 

 

30g/u

 

 

 

 

= 2 + 0 = 2 units

estimate = 2 units

(correction)

200mg/dL - 120mg/dL

- 3.5 units (active insulin)

= 2 - 3.5 = -1.5 units*

 

40mg/dL/u

 

 

 

 

*This negative number indicates that active insulin is sufficient to cover the correction that is needed. Thus, correction will be 0 units. Active insulin does not reduce the food portion of the estimate.

Default (factory) settings

Menu

Item

Default Setting

Limits

Increments

Warning

 

 

 

 

 

Limits

 

Bolus Wizard feature:

Off

 

 

 

 

 

 

 

 

 

 

Easy bolus:

On

 

 

 

 

 

 

 

 

 

 

Easy bolus step:

0.1 u/h

maximum bolus setting

 

 

 

 

 

 

 

 

Bolus

Dual/Square bolus:

Off

 

 

 

Menu:

 

 

 

 

 

Maximum bolus:

10.0 u/h

0 - 25 u (per single

 

 

 

 

bolus)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

BG reminder:

Off

0:00 - 5:00

0:30 (minutes)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Patterns:

Off

 

 

 

 

 

 

 

 

 

Basal

Maximum basal rate:

2.0 u/h

0.00 - 35.00 u/h

0.05u

 

 

 

 

 

 

Basal rate:

0.0 u/h

 

0.05u

 

Menu:

 

 

 

 

 

 

 

 

Temp basal type:

U/H

max basal rate setting

0.05U/H (or

(basal rate =

 

 

 

 

5%)

0.0)

 

 

 

 

 

 

154 Chapter: 11

Page 176
Image 176
Energizer 715, 515 manual Default factory settings