Is priming complete?
If you manually prime your pump with more than 30u insulin, this message appears. Press ESC, ACT to clear the message. If manual prime is complete, press ESC. If the manual prime is not complete, press and hold ACT until manual prime is complete.
Max delivery
This alarm alerts you when you have taken more insulin than expected based on maximum bolus and maximum basal rates.
Motor error
Insulin delivery has stopped. This alarm occurs if your pump detects a motor error.
No delivery
Insulin delivery has stopped. This alarm occurs if your pump detects a blockage.
No reservoir
The reservoir is not inserted correctly or no reservoir has been inserted.
IS PRIMING COMPLETE?
Disconnect from body See user guide
for instruction ESC, ACT to clear
3:25PMAX
DELIVERY
Exceeded 1 hour max delivery Check BG
ESC, ACT to clear
3:25P MOTOR
ERROR
Delivery stopped
Disconnect set
ESC, ACT to clear
11:17ANO
DELIVERY
Delivery stopped Change entire set Check BG
See user guide to troubleshoot ESC, ACT to clear
4:36PNO
RESERVOIR
Delivery Stopped
Change reservoir set
ESC, ACT to clear