Button error

Occurs if a button has been continually pressed for more than 3 minutes.

Check settings

When this alarm is active, you should check and/or reprogram your pump settings, including the time/date.

E (Error)

After receiving this alarm, note the error number then call the

Medtronic MiniMed 24-hour Product Help Line for assistance. An error alarm will show an “E” followed by two numbers. E-alarms cause all insulin delivery to stop, the pump resets, and all your settings are cleared.

Empty reservoir

There is no insulin in the reservoir. Change the reservoir immediately.

Failed batt test

The pump tests the voltage of each battery installed. This test ensures a battery with low voltage is not used. If the battery does not have enough voltage, this alarm will occur. The pump will not function and the battery must be replaced. (Always make sure that you install a NEW battery into the pump.)

10:05A BUTTON

ERROR

Button pressed for more than 3 min ESC, ACT to clear

8:35A CHECK

SETTINGS

Delivery stopped

Reprogram settings

ESC, ACT to clear

5:30AExx Settings cleared Reprogram settings

Call help-line for assistance

ESC, ACT to clear

XXindicates the error number

3:36P EMPTY

RESERVOIR

Delivery stopped

Change reservoir

ESC, ACT to clear

8:35A FAILED BATT TEST

Delivery stopped

Replace battery now

Use 1 AAA

ESC, ACT to clear

138 Chapter: 9

Page 160
Image 160
Energizer 715, 515 manual Error