138 Chapter: 9

Button error

Occurs if a button has been continually pressed for more than 3
minutes.

Check settings

When this alarm is active, you should check and/or reprogram your
pump settings, including the time/date.

E (Error)

After receiving this alarm, note the error number then call the
Medtronic MiniMed 24-hour Product Help Line for assistance. An error
alarm will show an ā€œEā€ followed by two numbers. E-alarms cause all
insulin delivery to stop, the pump resets, and all your settings are
cleared.

Empty reservoir

There is no insulin in the reservoir. Change the reservoir immediately.

Failed batt test

The pump tests the voltage of each battery installed. This test ensures
a battery with low voltage is not used. If the battery does not have
enough voltage, this alarm will occur. The pump will not function and the
battery must be replaced. (Always make sure that you install a NEW
battery into the pump.)
ERRORBUTTON
10:05A
Button pressed for
more than 3 min
ESC, ACT to clear
8:35A CHECK
SETTINGS
Delivery stopped
Reprogram settings
ESC, ACT to clear
Exx
XX indicates the
error number
5:30A
Settings cleared
Reprogram settings
Call help-line
for assistance
ESC, ACT to clear
3:36P EMPTY
RESERVOIR
Delivery stopped
Change reservoir
ESC, ACT to clear
8:35A FAILED
BATT TEST
Delivery stopped
Replace battery now
Use 1 AAA
ESC, ACT to clear