I dropped my pump
Take care to protect your pump from being dropped.
1Check that all connections are still tightly in place.
2Check the LCD, keypad and pump case for cracks or damage.
3Check infusion set, including the tubing connector and tubing for cracks or damage.
4Review the status screen, basal rates and other pump settings.
5Perform the Selftest procedure located in the UTILITIES MENU.
6Call the Medtronic MiniMed
I submerged my pump in water
Your pump is designed to resist accidental contact with water. Do not submerge in water during bathing, swimming, or other water activities.
1Pat the outside of the case until dry.
2Open the reservoir compartment and check the compartment and reservoir for water. If wet, dry it completely within ten (10) minutes of exposure to water. Exposure to liquids, including water or insulin can corrode the mechanism.
3Dry the reservoir
4Do not use hot air to dry your pump. This may damage your pump's internal electronics.
5Check the battery compartment and the
6Perform a selftest.