1Go to the ALERT TYPE screen.

Main > Utilities > Alarm > Alert Type

2Select your alert type and press ACT. That alert type is now active. Exit the menus.

Auto-off

ALERT TYPE

Beep Long

Beep Medium

Beep Short

Vibrate

The factory setting for this feature is off (set to 0 hours). This is a safety feature that stops insulin delivery after a defined time period (from 1 to 24-hours). If the pump detects that no buttons have been pressed for the selected amount of time, insulin delivery will stop and an alarm will sound. You may choose to program this feature into your pump based on the numbers of hours that you usually sleep at night. Discuss what uses and settings are best for you with your healthcare professional.

1Go to the AUTO OFF DURATION screen.

Main > Utilities > Alarm > Auto Off

2Set the number of hours you want to set and press

ACT.

AUTO OFF DURATION

0 Hours

(flashing)

NOTE - If you do not want to use the Auto Off feature, make sure the hour is set to zero (0).

3The screen will return to the ALARM MENU. The Auto Off feature is now set. Exit the menus.

Low resv alert (Low reservoir warning)

Allows you to program the pump to sound an alert before your reservoir is empty. You can select one of these warning types:

a specified number of units that remain in the reservoir

a specified maximum amount of time that remains before the reservoir will be empty

The factory setting for this feature is (20) insulin units.

1Go to the RESV WARNING TYPE screen.

Main > Utilities > Alarm > Low Resv Warning

2Select Insulin units or Time and press ACT.

RESV WARNING TYPE

Insulin Units

Time

WARNING: When the pump detects a low reservoir condition during a bolus or prime delivery, the alert will go off after the delivery is finished. Make sure to check the volume of your reservoir to ensure enough insulin is available.