1Go to the ALERT TYPE screen.
Main > Utilities > Alarm > Alert Type
2Select your alert type and press ACT. That alert type is now active. Exit the menus.
Auto-off
ALERT TYPE
Beep Long
Beep Medium
Beep Short
Vibrate
The factory setting for this feature is off (set to 0 hours). This is a safety feature that stops insulin delivery after a defined time period (from 1 to
1Go to the AUTO OFF DURATION screen.
Main > Utilities > Alarm > Auto Off
2Set the number of hours you want to set and press
ACT.
AUTO OFF DURATION
0 Hours
(flashing)
NOTE - If you do not want to use the Auto Off feature, make sure the hour is set to zero (0).
3The screen will return to the ALARM MENU. The Auto Off feature is now set. Exit the menus.
Low resv alert (Low reservoir warning)
Allows you to program the pump to sound an alert before your reservoir is empty. You can select one of these warning types:
➠a specified number of units that remain in the reservoir
➠a specified maximum amount of time that remains before the reservoir will be empty
The factory setting for this feature is (20) insulin units.
1Go to the RESV WARNING TYPE screen.
Main > Utilities > Alarm > Low Resv Warning
2Select Insulin units or Time and press ACT.
RESV WARNING TYPE
Insulin Units
Time
WARNING: When the pump detects a low reservoir condition during a bolus or prime delivery, the alert will go off after the delivery is finished. Make sure to check the volume of your reservoir to ensure enough insulin is available.