146 Chapter: 11
Meter value
The blood glucose measurement received from Paradigm Link meter. Appears in the ENTER BG screen during bolus
programming. Appears on the screen when the pump is idle at the HOME screen.
Normal bolus
Range 0.1 – 25.0 units of insulin (limited by maximum bolus setting).
Occlusion detection
When occlusion is detected, the “no delivery alarm” will occur. The occlusion alarm is triggered by an
average of 2.77 units of “missed” insulin. This table shows occlusion detection for 3 different situations
when using U100 insulin.
Percent temp basal
Default value: 100% of basal programming
Power supply
The pump is powered by a standard 1.5 V AAA alkaline battery (Energizer brand recommended).
Prime function
Fixed prime range0.1 – 25.0 units (limited by maximum bolus)
Manual prime limit: warning at 30 units, then at each 10 units thereafter.
Fill rate: ........ 1 to 5 units/second.
NOTE - Prime insulin is not counted in daily totals but is recorded separately in the prime history.
Prime history
Maximum records shown: 20 (manual and fixed)
Maximum ID entries expiration: range: warning limits:
3 12 minutes (mg/dL) 20 – 600
(mmol/L) 1.1 – 33.3 less than 20 or greater than 100 mg/dL
less than 1.1 or greater than 5.6 mmol/L
Rate Minimum time
before alarm
Typical time
before alarm
Maximum time
before alarm
bolus delivery (1.5 u/minute) 92 seconds 116 seconds 162 seconds
basal delivery (1.0 u/hour) 2.2 hours 3.09 hours 4.47 hours
basal delivery (0.05 u/hour) 37.4 hours 59.2 hours 87 hours