Pump maintenance 141

Chapter: 10

Pump maintenance
It is recommended that you read your warranty statement included with your pump for information
about what is covered during your warranty period.

Battery

The Paradigm pump uses a AAA battery. As a safety measure, Medtronic MiniMed has designed the
pump to only accept a NEW battery. If you insert a used battery, the “failed batt test” alarm may be
triggered. See “Install battery” on page 12 for instructions.
The use of cold batteries causes erratic pump behavior. To prevent this, do not use batteries that
have been in cold storage (i.e., in the refrigerator or your car in the winter). It takes several hours for
these batteries to warm to room temperature.
Certain features on the pump use a lot of battery power. Your battery will need to be replaced more
frequently if you use these features:

Storage

If you have to remove and store your pump, it is recommended that you store it with the battery in
place. Keep a record of your current basal rates. To preserve battery life, reset the basal rates to 0
(zero), turn off the remote and meter options, and set the Auto-off to dashes or zeroes.
Remote control Paradigm Link meter Backlight Vibrate alert type setting
CAUTION: It is recommended that you use a AAA Energizer battery. Do not use a
rechargeable or carbon zinc battery in the pump. Do not remove the battery
unless you are changing your battery (installing a NEW battery). Replace it
within five (5) minutes. If not replaced within five (5) minutes, the screen may
display an alarm message. Follow the instructions in the message and make
sure the time and date are set correctly. Install a new battery if the battery
was placed backwards in the pump.