Special - indicates a special feature is active or an alert condition(s)

 

exists. Special mode does not restrict any of the pump functions.

 

When the pump is in Special mode, an open circle appears at the

 

top of the screen and it will beep/vibrate periodically to remind

 

you of the condition. The conditions and features that will put the

 

pump in Special mode are:

Special Mode

Low Reservoir condition

Low Battery condition

Block feature is on

Dual or Square Wave bolus

Temporary basal

delivery

delivery

Basal pattern A or B is active

Attention - indicates insulin delivery has stopped. This can mean that the pump is in Suspend mode. It can also mean an alarm is active or an alarm condition exists that needs immediate attention for insulin delivery to resume. A solid circle appears at the top of the screen and the pump will beep periodically until

the either the pump is taken out of Suspend mode or the conditionAttention Mode is cleared. The screen will show text describing the condition that

put the pump in Attention mode. For example, if the reservoir is empty, “Empty Reservoir” will appear on the screen.

When the pump is in Attention mode, it will beep/vibrate periodically to remind you of the condition. The beep/vibrate frequency varies depending on the condition that put the pump in Attention mode. See “Alarm conditions” on page 137 for alarm conditions that will trigger the Attention mode. See “Stopping your pump” on page 39 to learn about Suspend mode.

18 Chapter: 2

Page 40
Image 40
Energizer 715, 515 manual Special Mode