Rewinding your pump

Before you continue, make sure the pump is NOT connected to your body.

WARNING: Make sure the infusion set is disconnected from your body before you rewind or prime the pump. Never insert the reservoir into the pump while the tubing is connected to your body. Doing so could result in an accidental infusion of insulin.

1If you removed your reservoir and are replacing it, go to the REWIND screen.

Main > Prime > Rewind

2In the REWIND screen, press ACT to start the rewind process. The REWINDING screen will appear while the pump rewinds.

3After the pump rewinds the PREPARING TO PRIME screen will appear.

NOTE - If you press ESC or the pump returns to the HOME screen the PRIMING STOPPED screen will appear which guides you back to the priming screens. Once you start a rewind, you cannot cancel it.

If you are practicing, do these steps:

REWIND

Disconnect from Body Press ACT to Rewind

PRIMING STOPPED

Continue Priming

Rewind Pump

Status Screen

ACT

ACT

REWINDING

 

 

Wait for notification

ESC PREPARING TO PRIME

Disconnect from Body Insert/Lock Reservoir

ACT Hold ACT to Prime

time out (HOME screen)

a.Do NOT insert the reservoir in your pump.

Make sure the shipping cap is installed in the reservoir compartment.

b.Continue with the manual prime instructions described in “Manual prime” on page 49. If you are not practicing, continue to “Inserting the reservoir in your pump” on page 48.

Starting on insulin 47

Page 69
Image 69
Energizer 515, 715 manual Rewinding your pump