Using the Bolus Wizard feature 71

NOTE - Pumps are sent from the factory with a BG target range of 90-140 mg/dL (5.0-7.8
mmol/L). If you adjust your BG targets outside of 90-140 mg/dL (5.0-7.8 mmol/L), the pump
screen displays a warning that the values are acceptable but outside the normal range.
1In the EDIT SETTINGS screen, select BG
Target and press ACT.
2Enter the low end of your BG target range
and press ACT. Then enter the high end of
your BG target range and press ACT.
NOTE - The start time for your first BG target range is midnight (12:00A or 00:00) and cannot
be changed.
3The screen will display one of these messages:
4To set just one BG Target: To set more than one BG Target:
a. Press ESC. a. Set the start time
for the next BG
target range and
press ACT.
b. Set the low and high
values for the next
BG target range and
press ACT.
c. Repeat steps a and b
to set more BG
target ranges.
d. Press ESC when you
are done.
->
- - -
first BG Target
Range
(flashing)
mg/dL
100 - 100
EDIT SETTINGS
BG Units: mg/dL
Sensitivity:
BG Target:
ACT TARGET RANGE 1
12:00A
_: _
(flashing)
press ESC
_ - _
SET START TIME 2
mg/dL _: _
(flashing) set, then press
ACT
- - -
SET START TIME 2*
7:30A
(flashing) set, then press
ACT
* Depending on how many
targets you set, this
number can be 2 through 8.
mg/dL
100 - 100
TARGET RANGE 2*