Energizer 715, 515 manual Bolus Wizard feature examples, Settings

Models: 715 515

1 188
Download 188 pages 55.73 Kb
Page 174
Image 174

Bolus Wizard feature examples

Settings:

Carb ratio: 30 grams/unit

Insulin Sensitivity: 40 mg/dL/unit

BG Target: 90-120 mg/dL

Active Insulin Time: 6 Hours

#1: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and does not enter blood glucose.

(food estimate)

60g

30g/u = 2 units

estimate = 2 units

#2: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a blood glucose of 200.

 

(food estimate)

 

 

60g

= 2 units

+

 

30g/u

 

 

 

 

 

= 2 + 2 = 4 units

estimate = 4 units

(correction)

200mg/dL - 120mg/dL= 2 units 40mg/dL/u

152 Chapter: 11