Chapter: 8

Utilities

Alarm review

You can review alarms in the ALARM HISTORY screen. This screen shows up to 36 past alarms and/or errors. You can also review the details for each alarm when you are in the ALARM HISTORY screen.

1Go to the ALARM HISTORY screen

Main > Utilities > Alarm > Alarm History

2Scroll through your past alarms.

3If you want to review the details for an alarm, continue to “Alarm details” below. Exit the menus if you are done.

ALARM HISTORY 11OCT 10:35P LoRes

08OCT 04:08P ChkSet 05OCT 11:35A LoBat

Alarm details

4In the ALARM HISTORY screen, select the alarm you want to review and press ACT. The details for that alarm will appear on the screen.

5Press ESC to return to the ALARM HISTORY screen. Select another alarm to review, or exit the menus if you are done.

LOW RESERVOIR 11OCT 10:35P

>24-hours remaining

0.25units remaining

Setting your alert type

You can select the type of alert your pump uses (for alarms, special conditions and programming). You can select a vibrate (silent) alert, or an audible beep alert. There are three beep types: long, medium and short tones. The factory setting for this feature is beep-medium.

The vibrate alert type is disabled if you use the block feature, and vibrate must be set again once block is turned off. If your alert type is set to vibrate and you get a LOW BATTERY alert, your pump will use the beep alert type instead to conserve battery power.

NOTE - Vibrate uses more battery power than the beep alert type and may shorten battery life.

Utilities 109

Page 131