User settings
The user settings function allows you to save, restore, and clear all pump settings. You can also view a listing of the dates and times of all recent user settings operations you have done. The Save Settings feature lets you keep a set of pump settings that you can restore to your pump if it is cleared or you need to go back to these settings for any reason.
When you clear your pump, the pump settings are restored to the factory defaults, and you must either use Restore Settings, if you have saved a set of pump settings, or reprogram all your settings before you can use your pump again. The pump does not clear the internal pump memory.
WARNING: Do not clear your pump settings while it is connected to your body.
CAUTION: Do not clear your pump settings unless directed by your healthcare professional or a Medtronic MiniMed representative. If you clear your pump settings, it will be necessary to reprogram all your personal pump settings as directed by your healthcare professional. Additionally, you will have to rewind your pump.
Saving pump settings
Do these steps to save your current pump settings:
1Go to the Utilities screen and highlight User Settings. Then, hold
down and press ACT.
Main > Utilities > User Settings
2The USER SETTINGS screen is displayed with Save Settings highlighted. Press ACT.
UTILITIES MENU
Selftest
User Settings
Language
ACT
USER SETTINGS
Save Settings
Clear Settings
History
ACT