User settings

The user settings function allows you to save, restore, and clear all pump settings. You can also view a listing of the dates and times of all recent user settings operations you have done. The Save Settings feature lets you keep a set of pump settings that you can restore to your pump if it is cleared or you need to go back to these settings for any reason.

When you clear your pump, the pump settings are restored to the factory defaults, and you must either use Restore Settings, if you have saved a set of pump settings, or reprogram all your settings before you can use your pump again. The pump does not clear the internal pump memory.

WARNING: Do not clear your pump settings while it is connected to your body.

CAUTION: Do not clear your pump settings unless directed by your healthcare professional or a Medtronic MiniMed representative. If you clear your pump settings, it will be necessary to reprogram all your personal pump settings as directed by your healthcare professional. Additionally, you will have to rewind your pump.

Saving pump settings

Do these steps to save your current pump settings:

1Go to the Utilities screen and highlight User Settings. Then, hold

down and press ACT.

Main > Utilities > User Settings

2The USER SETTINGS screen is displayed with Save Settings highlighted. Press ACT.

UTILITIES MENU

Selftest

User Settings

Language

ACT

USER SETTINGS

Save Settings

Clear Settings

History

ACT

Utilities 123

Page 145
Image 145
Energizer 515, 715 manual User settings, Saving pump settings