WARNING: Your pump will not accurately deliver insulin if there are air bubbles in the infusion set. Take care to remove any air bubbles during the manual prime.

IS PRIMING COMPLETE?

Disconnect from body See user guide

for instruction. ESC, ACT to clear

If your manual prime uses more than 30U insulin, the pump will ask you, “is priming complete?” If you get this message, do these steps:

a.Make sure you are not connected to the pump.

b.Read the message on the screen, then press ESC, ACT to clear.

a.If your manual prime is complete (you see droplets on the tip of the infusion set needle), press ESC and continue to step 5. If your manual prime is not complete, press and hold ACT until priming is complete then continue to step 5.

5Press ESC. Your manual prime is complete.

6You can now insert the infusion set into your body as described in the next section.

Insert the infusion set

WARNING: While the infusion set is connected to your body, do not unscrew and retighten the tubing connector on the reservoir.

After you complete all of the following, you will be ready to insert the infusion set into your body:

fill your reservoir

rewind your pump

and prime the pump (fill the infusion set with insulin)

It is important that you change your infusion set every 2-3 days. Medtronic MiniMed offers a number of different infusion sets for your pump. Instructions for the Quick-set®begin on the next page as an example. Always refer to the instructions that shipped with your infusion set. After your infusion set is inserted, continue to “Fixed prime” on page 55.