1 174
Download 174 pages 9.17 Kb
Page 11
Image 11
Specifications

Chapter 1: General Information

Specifications

The specifications are the performance standards against which the product is tested.

Maximum State Speed (selectable) Minimum Master to Master Clock Time* Threshold Accuracy

Setup/Hold Time*

Single Clock, Single Edge

200MHz 5.000 ns

±(65 mV + 1.5% of threshold setting)

4.5/-2.0 ns through -2.0/4.5 ns, adjustable in 100 ps increments

Single Clock, Multiple Edges

5.0/-2.0 ns through -1.5/4.5 ns, adjustable in 100 ps

 

increments

Multiple Clocks, Multiple Edges

5.0/-2.0 ns through -1.5/4.5 ns, adjustable in 100 ps

 

increments

* Specified for an input signal VH = 0.9 V, VL = 1.7 V, slew rate = 1 V/ns, and threshold = 1.3 V.

Characteristics

These characteristics are not specifications, but are included as additional information.

 

Full Channel

Half Channel

Maximum State Clock Rate

150 MHz

not applicable

Maximum Conventional Timing Rate

250 MHz

500 MHz

Memory Depth (1680A, or 1690A-series)

512 K

1024 K

Memory Depth (1680AD or 1690AD-series)

2048 K

4196 K

Channel Count

 

 

1680A,AD or 1690A,AD

136

68

1681A,AD or 1691A,AD

102

51

1682A,AD or 1692A,AD

68

34

1683A,AD or 1693A,AD

34

17