Agilent Technologies 1680 To test the multiple-clock state acquisition, Set up the equipment

Models: 1690 1680

1 174
Download 174 pages 9.17 Kb
Page 48
Image 48
To test the multiple-clock state acquisition

Chapter 3: Testing Performance

To test the multiple-clock state acquisition

To test the multiple-clock state acquisition

Testing the multiple-clock, state acquisition verifies the performance of the following specifications:

Minimum master-to-master clock time.

Maximum state acquisition speed.

Setup/Hold time for multiple-clock, state acquisition.

This test checks two combinations of data using multiple clocks at two selected setup/hold times.

Equipment Required

Equipment

Critical Specifications

Recommended Model/Part

 

 

 

Pulse Generator

200 MHz 3.0 ns pulse width,

8133A option 003

 

< 600 ps rise time

 

Digitizing Oscilloscope

6 GHz bandwidth, < 58 ps rise time

54750A w/ 54751A

Adapter

SMA(m)-BNC(f)

1250-1200

SMA Coax Cable (Qty 3)

18 GHz bandwidth

8120-4948

Coupler

BNC(m)(m)

1250-0216

BNC Test Connector,

 

 

6x2 (Qty 4)

 

 

 

 

 

Set up the equipment

1If you have not already done so, do the following procedures:

“To set up the test equipment and the logic analyzer” on page 23. “To set up the logic analyzer for the state mode tests” on page 33.

2Increase the pulse generator channel 2 width to 3.000 ns.

48

Page 48
Image 48
Agilent Technologies 1680, 1690 manual To test the multiple-clock state acquisition, Set up the equipment