Agilent Technologies 1690 manual 1680A,AD-series disassembly/assembly

Models: 1690 1680

1 174
Download 174 pages 9.17 Kb
Page 114
Image 114
1680A,AD-series disassembly/assembly

Chapter 6: Replacing Assemblies

1680A,AD-series disassembly/assembly

1680A,AD-series disassembly/assembly

Prepare the instrument for disassembly

Do this procedure before doing any disassembly procedure on the instrument.

1Close the Agilent Logic Analyzer application software.

2Gracefully shut down the operating system and remove power when shutdown is complete.

3Remove the power cord.

4Move the instrument to a static safe work environment.

To remove the chassis from the sleeve

Before disassembling the instrument, it must be turned off and placed in a static safe work environment. If you haven't already done so, do the previous procedure "Prepare the instrument for disassembly."

1Using a Torx T-15 screwdriver, remove the two screws that secure the handle to the side of the instrument and lift off the handle.

2Using a Torx T-10 screwdriver, remove five screws that secure the logic analyzer cables to the rear panel of the logic analyzer.

3Disconnect the logic analyzer cables from the rear panel. Remove the logic analyzer cables (and spacers, if installed) from the logic analyzer.

4Using a Torx T-15 screwdriver, remove the screws connecting the four rear feet to the instrument, one screw per foot. Remove each foot from the rear panel.

5Using a Torx T-10 screwdriver, remove eleven screws that secure the sleeve to the chassis.

6With the logic analyzer upright, slide the chassis out of the sleeve.

114

Page 114
Image 114
Agilent Technologies 1690 manual 1680A,AD-series disassembly/assembly, Prepare the instrument for disassembly