then select [Enter] to display the setup time (∆ Time(1)-(2)).

Chapter 3: Testing Performance

To test the single-clock, single-edge, state acquisition

then select [Enter] to display the setup time (Time(1)-(2)).

cAdjust the pulse generator channel 1 Delay until the pulses are aligned according to the setup time of the setup/hold combination selected, +0.0 ps or –100 ps.

9Select the clock to be tested:

The following clock configurations will be used in steps 9, 10 and 11.

aIn the Analyzer Setup dialog, click the Sampling tab.

bIn the Sampling tab, click the Master button for the first clock to be tested (Clk 1) and select Falling Edge.

45