Connect and configure the logic analyzer

Chapter 3: Testing Performance

To test the threshold accuracy

Connect and configure the logic analyzer

1Using the 17-by-2 test connector, BNC cable, and probe tip assembly, connect the data and clock channels of Pod 1 to the free side of the BNC Tee.

2Configure the logic analyzer:

a Click the Manual background Bus/Signal Setup icon. The Analyzer Setup dialog opens. b In the Buses/Signals tab, click Delete All at the bottom of the dialog.

c Using the mouse, activate all Pod 1 channels. Assign channels to bus/ signal name My Bus 1.

d Scroll the channel assignments to the left. Assign the clock/data channel for the Pod 1 (that is, C1) to My Bus 1.

3Activate the DC source output.